A systematic, computerized database search was conducted using MEDLINE (PubMed) and the Cochrane Library for studies from 2004 until the search date, January 2016, using the medical subject heading (MeSH) keywords “navigation”, “dental implants”, and “accuracy”. A total of 37 papers were identified. Inclusion criteria were: a sample size of five or more implants, in vivo studies, and placement after ablative tumor surgery. Exclusion criteria were the following: a sample size of less than five implants, in vitro studies, and ex vivo studies. Four articles met aforementioned criteria and were included in this review.
Mean deviations were less than 1.4 mm in any direction and 6.5 degrees from the long axis of the implant. The table below describes the details of the selected studies.
Navigation | Platform | Error | (mm) | Apex | Error | (mm) | Angle | Error | (degrees) | |||
Study | Implants | System | Direction | Mean | SD | Max | Mean | SD | Max | Mean | SD | Max |
Elian et al (2008) New York, NY |
14 | DenX Advanced¹ |
Global |
0.89 |
0.53 |
1.96 |
0.96 |
0.50 |
1.99 |
3.78 |
2.76 |
9.86 |
Wittwer et al (2007) Vienna, Austria |
32
|
VISIT²
|
Buccal Lingual
|
1.00 0.70
|
0.50 0.30
|
2.00 1.20
|
0.60 0.70
|
0.20 0.30
|
0.90 1.00
|
|||
|
32 | Treon³ |
Buccal Lingual |
1.00 1.20 |
0.50 0.80 |
2.40 3.40 |
0.80 0.70 |
0.60 0.50 |
2.00 1.60 |
|||
Wittwer et al (2006) Vienna, Austria |
78 |
Treon³ |
Global |
1.10 |
0.70 |
3.40 |
0.80 |
0.60 |
2.00 |
|||
Wagner et al (2003) Vienna, Austria |
32 | VISIT² | Buccal Lingual |
0.80 1.00 |
0.30 0.50 |
2.10 2.60 |
1.10 1.30 |
0.90 0.90 |
3.40 3.50 |
6.40 |
3.65 |
17.40 |
¹Dental Systems, Moshav Ora, Israel | ||||||||||||
²University of Vienna, General Hospital, Vienna, Austria | ||||||||||||
³Medtronic, Minnesota, MN |
Dynamic navigation allows the accurate real-time transfer of the virtual surgical plan to the patient even in difficult implantologic situations. Results were consistent among different surgeons with varying degrees of familiarity with the protocol operating in geographically distinct areas around the world. Dynamic navigation uses a time-efficient workflow to place implants with similar placement error to static navigation. In dynamic navigation systems, both engineering (intrinsic) and human usability factors (extrinsic) can be visualized for accuracy. Dynamic navigation is used during dental implant placement to: (1) avoid critical anatomic structures, (2) accurately place multiple implants with proper spacing and angulation, (3) place single implants in exact locations when access is limited and esthetic needs are high, and (4) minimize flap mobilization to achieve minimally invasive surgery.
Block MS, Emery RW, Static or Dynamic Navigation for Implant Placement – Choosing the method of guidance, J Oral Maxillofac Surg 74:269-277, 2016.
Wittwer G, Adeyemo WL, Prospective randomized clinical comparison of 2 dental implant navigation systems, Int J Oral Maxillofac Implants. 2007 Sep-Oct;22(5):785-90.